| Sign In | Join Free | My infospaceinc.com |
|
Brand Name : KACISE
Model Number : KGD-MQ-516
Certification : CE
Place of Origin : CHINA
MOQ : 1PCS
Payment Terms : L/C,D/A,D/P,T/T,Western Union,MoneyGram
Supply Ability : 1000 Piece/Pieces per Week negotiable
Delivery Time : 5-8 work days
Packaging Details : each unit has individual box and all boxes are packed in standard packages or customers requests available
Measured Components : CO/CO₂/CH₄/H₂/O₂/C₂H₂/C₂H₄/CₙHₘ*/Calorific Value
Measurement Principle : CO/CO₂/CH₄/C₂H₂/C₂H₄/CₙHₘ: NDIR; H₂: TCD; O₂: ECD
Measurement Range : CO/CO₂/CH₄/H₂: (0 - 100)%; O₂: (0 - 25)%; C₂H₂/C₂H₄/CₙHₘ: (0 - 10)% (Optional)
Linear Error : CO/CO₂/CH₄/C₂H₂/C₂H₄/CₙHₘ: ±1%F.S.; O₂/H₂: ±2%F.S.
Resolution : 0.0001
Repeatability : ≤ 1%
The On-line Coal Gas Analyzer KGD-MQ-516 employs non-dispersive infrared (NDIR) gas analysis technology with independent intellectual property rights, long-life electrochemical sensing technology, and MEMS-based thermal conductivity technology. It can simultaneously measure the calorific values of coal gas and biogas online, as well as the volume concentrations of gases including CO, CO₂, CH₄, H₂, O₂, and CₙHₘ.
| Measured Components | CO/CO₂/CH₄/H₂/O₂/C₂H₂/C₂H₄/CₙHₘ*/Calorific Value |
|---|---|
| Measurement Principle | CO/CO₂/CH₄/C₂H₂/C₂H₄/CₙHₘ: NDIR; H₂: TCD; O₂: ECD |
| Measurement Range | CO/CO₂/CH₄/H₂: (0-100)%; O₂: (0-25)%; C₂H₂/C₂H₄/CₙHₘ: (0-10)% (Optional) |
| Linear Error | CO/CO₂/CH₄/C₂H₂/C₂H₄/CₙHₘ: ±1%F.S.; O₂/H₂: ±2%F.S. |
| Resolution | 0.01% |
| Repeatability | ≤ 1% |
| Response Time | T90 < 15s (NDIR) |
| Optimal Flow Rate | (0.7-1.2) L/min |
| Inlet Pressure | (2-50) kPa |
| Gas Requirements | Dust-free, water-free, oil-free |
| Communication | RS-485/RS-232, (4-20)mA |
| Power Supply | 100-240VAC, 100W |
| Display | LCD Display |
|
|
Online Coal Gas Analyzer with Multi-Component Measurement High Accuracy and Automated Pre-Treatment for Biogas Analysis Images |